1200 Liter Large Steam Sterilizer With Safety Valves In Jacket And
Chamber
Specification
Model No | MD-1.2Y | |
Chamber dimensions | mm | 680*1180*1500 |
Overall dimensions | mm | 2200*2000*1910 |
Inner capacity | L | 1200 |
Door operation | AUTO | Horizontal single sliding |
Setting range time | min | 0~99 |
Setting range temperature | ℃ | 110~135 |
Controller | Panel | Touch screen |
System | PLC | |
Display | 5.7" Touch Screen | |
Version | Chinese/English/Spanish/Thai | |
Output interface | RS-232 | |
Vacuum | Drawing | Vacuum pump |
Printer | Paper | Heat sensitive type with auto rolling back |
Accessories | Trolley | 1 set of floor-Loading carriage |
Printer paper | 2 rolls | |
Manual | 1 copy of operation and maintenance manual |
Program Parameter
Cycle name | Further details | Application |
Instruments/Tools | 3 times Vacuum, expose temperature 132℃, expose time 8 min, drying time 5 min. | Solid materials to be sterilized: |
Clothing | 3 times Vacuum, expose temperature 121℃, expose time 15 min, drying time 25 min. | Materials to be sterilized from which air can’t be easily displaced |
User Programmable | All preset parameters can be reset; Multiple customized programs can be set and stored by user. | Customized |
Test programs | ||
B-D Test | 3 times Vacuum, expose temperature 134℃, expose time 4 min, drying time 3 min. | Test program |
Vacuum Leakage Test | Test the chamber connecting parts’ tightness effect. It’s important to ensure the sterilization effect. | Test program |
Description of sterilization cycle
Pre-Vacuum Cycle - pre-conditioning of load during air removal using programmable
ramped
positive and negative pressure pulsing to remove entrapped air.
Programmable ramped vacuum
drying.
Micro Isolator Cycle - pre-vacuum cycle using programmable, ramped, positive and negative
pressure pulsing to reduce internal and external crazing of animal
housing by allowing pressure stabilization to product items being
processed.
Test Cycles Leak Test - programmable vacuum leak test to verify door seal, sterilizer
chambe
r and piping integrity through monitoring of acceptable vacuum loss
over selected time.
Autoclave air removal test - pre-programmed vacuum cycle used to verify removal of residual air
from
chamber and load, and ensure steam penetration into the test load.
Bio-Waste Cycle - designed for pre-sterilization conditioning through positive and
negative pressurization to ensure air removal from mixed products
in containers. Programmable ramped steam pressurization is used to
maximize heat penetration of mixed laboratory biohazard waste.
F0 Temperature Control Cycle - for the sterilization of heat sensitive materials using
time-at-temperature calculations beginning at a predetermined point
of the conditioning phase. This reduces the product exposure to
higher temperatures normally found in standard sterilization
cycles.
Gravity Cycle - for non-air retentive products; utilizes positive steam pressure
with vacuum assist air removal for conditioning of load. The end of
cycle exhaust is programmable for dry and non-dry phases
Liquid Cycles Liquid Cycle - cycle used for liquids in vented borosilicate glass or metal
containers. Positive steam pressure with vacuum assisted air
removal with programmable ramped heating and exhaust to ambient
with optional cooling phase.
Liquid, Air Cool Cycle - cycle used for liquids in vented borosilicate glass or metal
containers
. Positive steam pressure with vacuum assisted air removal with
programmable ramped
heating and introduction of filtered air during exhuast to speed
cooling.
Pre-Vacuum Liquid Cycle - pre-conditioning of load using programmable, ramped, positive
and negative pressure pulsing to assist in removal of entrapped
air. Chamber temperatures
at negative pressures are controlled to prevent boiling of product.
Controlled ramped exhaust
to ambient with optional cooling phase.
Effluent - for the processing of items in BSL3 and BSL4 laboratories. The
cycle uses a heated,
jacketed, 0.2 micron hydrophobic filter to contain biohazard
contents in sterilizer chamber
while allowing exhaust of sterile gasses.
Advantages
Reliability - Industrial grade valves and high temperature gaskets prevent
otherwise costly
and common service downtime. The vessel is tested and certified to
exceed national pressure
vessel standards and carries a manufacturer’s warranty.
Safety - Operator and technician safety is ensured through temperature and
pressure monitoring,
relief valves, safety interlocks overrides.
Serviceability - Reduce costly maintenance downtime with easy to access consumable
and
expendable components.
Simplicity - The intuitive touch screen interface provides users with clearly
identifiable selections
to begin cycles.
Long Life Door Gasket - Medpharm high temperature, air actuated door gaskets are seated
into a machined groove that provides a close tolerance creating a
less strenuous environment,
extending door gasket life.
Brochure load down from here
ZHANGJIAGANG MEDPHARM MACHINERY LTD.
MEDPHARM Company team is committed to providing quality sterilization products, service, parts and
maintenance training at significant savings. From customized parts packages to turnkey equipment
acquisition and installation, our innovative sterilization solutions are designed to meet our customers' needs.
We understand that untimely breakdowns are costly and can be detrimental to your everyday operations.
Our service team is compromised of expert technicians that respond quickly to all of your service needs.
We offer on-site service support including inservices, installation and deinstallation of your equipment. From
breakdown repairs to comprehensive maintenance programs, we have an innovative service solution for you.
We will service on-Site Repairs.Our staff is equipped with the resources required to provide you with
on-site, on-time repairs in compliance with manufacturer specifications. We also provide installation,
deinstallation, and inservices on most makes and models of sterilization equipment. Near or far we can
provide you with service support that you need.
Our reconditioning background has taught us of the importance of having quality parts on hand at all times.
We have a large inventory of individual repair parts, as well as preventive maintenance kits, and consumables
in stock and ready to ship today.
With our expert technical support staff, it is not always necessary to have part numbers to place an order.
We will help you identify the quality parts that you need.